Name: | Ultra High Vacuum Preparation and Analytical system - Photoelectron Spectroscopy SPECS Phoibos 150 |
Location: | CEITEC Brno University of Technology |
Instrument description: | The instrument is equipped with the Al Kα and Mg Kα X-ray sources which provide photons with 1486.6eV and 1254.6eV energy respectively and He I/II UV source providing photons with energies of 21.22eV and 40.81eV respectively. The photoelectrons are analyzed in a hemispherical detector in which lens voltages direct the electron onto a two-dimensional (energy, momentum) multi-channel plate allowing for angular-resolved photoelectron spectroscopy (ARPES) experiments. For sample handling there is a 4-axis manipulator (X, Y, Z, Polar rotation). The manipulator allows controlled sample (electron beam) heating up to 800°C (peak short flashes up to 1200°C) as well as liquid nitrogen cooling. The system allows for neutralization of non-conductive samples by means of an electron flood gun. The standard sample holder is compatible with samples of 10mm diameter. There is no general restriction for the sample material (except for Zn) however, compatibility of volatile and/or organic materials should be discussed and agreed in advance. |
Research group: | CF: CEITEC Nano |
If you would like to use this equipment contact the administrator: |
---|